Your best just got better.
XOS’s next generation of analyzer provides unparalleled precision with enhanced UX, improved data management, and greater ease of use.
VIDEO
LIMS integration for data management and transfer
Custom sample presets to save data entry time and minimize data entry error on common samples
Bar code reader auto fills sample name to reduce data entry time
Storage capacity for more than 50,000 measurement results
Supports up to 30 calibration curves
USB connectivity in front and back for connecting to printer, keyboard, mouse, and memory stick
Supports USB and network printers
Large, easy-to-remove side panels for easy serviceability
Advanced error reporting and diagnostics
Optional Autosampler to maximize productivity: 8-cup 43mm standard cup or 10-cup Accucells model
Product Details
Applications
Total sulfur analysis from ultra-low sulfur fuels to crudes
Total chlorine analysis from aqueous solutions and aromatic products to heavy fuels and crudes
For use in refinery labs, pipeline terminals, additive plants and inspection laboratories
Features
Sulfur
LOD: 0.4 mg/kg (ppm) at 300s, 0.28 mg/kg (ppm) at 600s*
Dynamic Range: 0.4 mg/kg (ppm) to 5 wt%
Chlorine
LOD: 0.3 mg/kg (ppm) at 300s, 0.21 mg/kg (ppm) at 600s*
Dynamic Range: 0.3 mg/kg (ppm) to 3000 mg/kg (ppm)
Automatic sulfur correction saves time and improves accuracy and precision on high sulfur samples
Easy to use
Intuitive 10-inch touch screen
Just plug in and measure
Measurement time: 10-999 s
Low maintenance: no gasses, or heating elements
Traditional 43 mm XRF sample cups or XOS Accucells – decided at time of order
Small footprint
Specifications
Test Method
ASTM D2622, D7039, D7536, D4929 ISO 20884, and SH/T 0842
Dimensions
42 cm (h) x 40 cm (w) x 54 cm (d) 16.5 in (h) x 15.8 in (w) x 21 in (d)
Power
100-120 VAC, 47-63 HZ at 5.0 Amps 200-240 VAC, 47-63 HZ at 2.5 Amps
Minimum Sample Cup Volume
Traditional - 5 ml, Accucell – 1 mL
Ambient Temperature Requirements
5-40° C (40-104° F)
Excitation Source
75 W air-cooled
* All qualifications herein are subject to user guide specifications.
* Longer cycle time increases counts and lower LOD, but sample condition over time needs to be considered.