- Analyzers
- Optics & Sources
- Technologies
- Support
- About
Clora 2XP delivers twice the precision for total chlorine analysis in liquid hydrocarbons such as aromatics, distillates, heavy fuels and crude oils, as well as aqueous solutions. Compliant with ASTM D7536 and D4929 methodology, Clora 2XP is ideal for testing related to catalyst poisoning in reformers, and sites with catalytic crackers and hydrocrackers. In addition, its automatic sulfur correction is perfect for high sulfur and low chlorine applications, such as crude oil and VGO. Powered by MWDXRF, Clora 2XP does not require gasses or high temperature processes, equating to easy operation and minimal maintenance requirements.
Corrosion at petroleum refineries is a critical problem all over the world. In the United States alone, NACE International, a professional association dedicated to reducing the economic impact of corrosion, has shown that $3.7 billion in direct costs are realized annually from maintenance, vessel expenditures, and fouling as a result of corrosion.1 This huge figure does not take into account the estimated loss of as much as $12 billion in profit due to decreased capacity, unit outages, and premature turnarounds that result from corrosion.2
Since its launch in 2007, Clora® has been widely adopted by refineries and test labs for monitoring chlorine in petroleum. With over 200 units in the field, customers trust Clora for testing products from crude oil to naphtha cuts and vacuum gas oil (VGO). Implementing corrosion mitigation strategies is an important part of ensuring safe refinery operation and maximizing profitability. Clora has become a critical part of identifying potential corrosion events and monitoring the effectiveness of these mitigation strategies.
As they continue to battle the huge costs of corrosion, leading refiners have found that continuing to lower the chloride content of feedstocks and process streams has significantly extended turnaround times and lowered the costs of corrosion. XOS has responded to the industry’s need by introducing Clora 2XP analyzer, delivering enhanced detection down to 0.07 ppm in hydrocarbon-based samples. This enables analysis at roughly 0.25 ppm for total chlorine. The increased capability enables the user to better understand and manage the most demanding samples like desalted crude and vacuum gas oil.
Various hydrocarbon samples were prepared and analyzed following a modified ASTM D7536 procedure. The sample types analyzed were:
The samples were pipetted into typical XRF sample cups and analyzed for 600 seconds. The standard deviation and average was then calculated based on the results of the sample runs, shown below in Table 1.
Table 1: Clora 2XP Total Chlorine Analysis Results | |||||||
---|---|---|---|---|---|---|---|
Gasoline | VGO | Naphtha | Mineral Oil | ||||
Repeats | Results | Repeats | Results | Repeats | Results | Repeats | Results |
#1 | 0.29 | #1 | 1.41 | #1 | 0.58 | #1 | 0.30 |
#2 | 0.31 | #2 | 1.42 | #2 | 0.54 | #2 | 0.33 |
#3 | 0.30 | #3 | 1.44 | #3 | 0.40 | #3 | 0.31 |
#4 | 0.33 | #4 | 1.36 | #4 | 0.52 | #4 | 0.31 |
#5 | 0.36 | #5 | 1.43 | #5 | 0.49 | #5 | 0.30 |
#6 | 0.40 | #6 | 1.35 | #6 | 0.55 | #6 | 0.27 |
#7 | 0.36 | #7 | 1.44 | #7 | 0.48 | #7 | 0.23 |
#8 | 0.32 | #8 | 1.47 | #8 | 0.47 | #8 | 0.34 |
#9 | 0.32 | #9 | 1.39 | #9 | 0.50 | #9 | 0.32 |
#10 | 0.31 | #10 | 1.46 | #10 | 0.51 | #10 | 0.34 |
Mean | 0.327 | Mean | 1.417 | Mean | 0.51 | Mean | 0.305 |
Standard Deviation | 0.032 | Standard Deviation | 0.040 | Standard Deviation | 0.05 | Standard Deviation | 0.035 |
All values in parts per million (ppm) |
The results of this experiment demonstrate the capability of Clora 2XP to precisely and repeatably measure sub-ppm chlorine concentrations in various hydrocarbon samples. This simple measurement requires just a few minutes to complete and requires no consumable gases or chemicals. Effectively monitoring chloride concentration in crude feedstocks and process streams is a critical part of any corrosion mitigation strategy. With the capability to quantify chlorine at sub-ppm levels in hydrocarbon samples, refiners will be able to reduce the multi-billion dollar costs of corrosion.